[18p-P01-17]Development and performance evaluation of a 1024-channel 2D MPPC-based PC-CT system for dynamical imaging

〇(M1)Minori Oshima1, Makoto Arimoto1, Takahiro Tomoda1, Fitri Lucyana1, Yu Furuta1, Jun Kataoka2, Ryotaro Minagawa2, Shinsuke Terazawa3, Satoshi Shiota3 (1.Kanazawa Univ., 2.Waseda Univ., 3.Proterial Ltd.)

Keywords:

Photon-Counting CT,Multi-Pixel Photon Counter,dynamical imaging

We are developing a photon-counting CT (PC-CT) system using a high-speed scintillator (YGAG:Ce) and a photodetector element, MPPC (Multi-Pixel Photon Counter). In this study, we developed a PC-CT system using a 64 × 16 pixel (1024 channels) 2D array detector for in vivo dynamical imaging of mice. since CT imaging requires uniformity of performance, we used a correction function to suppress the variation in response between pixels to achieve uniform performance. The system was developed using a correction function that suppresses the variation in response between pixels.