Presentation Information
[8p-C310-13]NV Magnetometer with Flux Concentrators for human MEG
〇(M2)Yudai Baba1, Tingyu Zhu1, Naota Sekiguchi2, Chikara Shinei3,4, Masashi Miyakawa4, Takashi Taniguchi4, Tokuyuki Teraji4, Hiroshi Abe5, Shinobu Onoda5, Takeshi Ohshima5,6, Masaki Sekino1 (1.The Univ. of Tokyo, 2.Science Tokyo, 3.Univ. of Tsukuba, 4.NIMS, 5.QST, 6.Tohoku Univ.)
Keywords:
NV magnetometer,MEG,flux concentrator
To realize human magnetoencephalography (MEG) measurements, we evaluated an NV magnetometer integrated with flux concentrators (FCs). Dry-phantom measurements demonstrated a magnetic field enhancement factor of approximately 60, and the sensitivity was estimated to be approximately 7 pT/Hz1/2. Assuming a brain magnetic field of 200 fT, approximately 5,000 signal averages are required for MEG measurements.