Presentation Information

[9p-P08-3]Study on neutron response in transparent composite scintillators

〇(D)Yuya Oshima1, Yutaro Saito1, Kenichi Watanabe1 (1.Kyushu Univ.)

Keywords:

neutron detector,scintillator,Li glass

To evaluate the neutron dose in BNCT, we fabricated transparent composite scintillators using Li glass powder and UV-curable resin, and the formation mechanism of the energy peak from the 6Li(n,t)α reaction was investigated. It was also shown that the energy peak position and width vary with the particle size of the Li glass, highlighting the importance of appropriate particle size selection.