Presentation Information
[9p-E204-5]Development of Effective Triplet Exciton Capture System for α-Particle Plastic Scinitllator
〇(DC)Kota Inage1, Marina Kanno2, Mitsugu Sohma3, Jun Akedo3, Junichi Kaneko2,4, Mikio Higuchi4, Yuki Morishita2, Yuichi Kitagawa4,5 (1.Grad. Sch. Chem. Sci. Eng., Hokkaido Univ., 2.JAEA, 3.AIST, 4.Fac. Eng., Hokkaido Univ., 5.WPI-ICReDD, Hokkaido Univ.)
Keywords:
plastic scintillator,alpha-particle,europium
Plastic scintillators are promising candidates for detecting alpha-particles owing to their high processability and transparensy. However, loaded luminophores in the polymer matrix have two drawbacks: (1) they cannot utilize triplet excitons in the polymer matrix and (2) they exhibit concentration quenching, resulting in less exciton capture efficiency from the polymer matrix. In this work, we developed a highly sensitive plastic scintillator incorporating a Eu(III) complex at high concentration levels, which exhibited tens of times greater scintillation than international standard plastic scintillator (BC-400, Saint-Gobain).