[1A02]Characterization of TlBr Gamma Detector Based on Electrical Charge and Cherenkov Light Analysis
*Boyu Feng1, Moh HAMDAN1, Keitaro Hitomi2, Mizuki Uenomachi3, Mitsuhiro Nogami2, Hidenori Toyokawa4, Kenji Shimazoe1 (1. UTokyo, 2. Tohoku Univ., 3. Tokyo Tech, 4. JASRI)
Keywords:
TlBr detector,Cherenkov Light,Gamma-ray detector
In this study, we developed a Thallium Bromide (TlBr) semiconductor detector, featuring a pixelated anode and a planar cathode, for room-temperature gamma-ray detection. The detector used Tl electrodes and was connected to a charge-sensitive preamplifier via Au wire for the electrical readout system. A silicon photomultiplier is used for an optical readout system to detect Cherenkov light. We achieved an enhanced energy resolution from 4.5% to 1.96% FWHM at 662 keV using depth of interaction correction. Time resolution assessments indicated potential performance in the picosecond range, showcasing TlBr's capability for precise energy resolution and timing.